The Impact of Internal Electric Fields on Accurate Diamond Magnetometry
arXiv:2611.03984v1 Announce Type: new Abstract: Nitrogen-vacancy (NV) center spin ensembles are a promising platform for accurate vector magnetometry, as they offer measurements referenced to a spin Hamiltonian in an anisotropic crystal with native vector sensing capabilities. However, charged dopants within the diamond produce random internal electric fields (RIEFs) that shift NV spin-transition frequencies, and their impact on the accuracy of NV magnetometry has not been investigated. Here, we characterize RIEFs' effect on NV spin-transition spectra and propose an analytical fitting function to improve accuracy in their presence. We build a Monte Carlo RIEF model that reproduces the
…