Modeling Valve-Closure Pressure Spikes in Miniature Fluid Systems

When a miniature liquid-handling system behaves well at steady state but produces a sharp pressure pulse at every valve transition, the valve is not necessarily defective. The transient can be the predictable result of stopping a moving liquid column faster than the rest of the fluid path can absorb its momentum. This matters in IVD analyzers, reagent dispensers, environmental instruments, and laboratory-automation modules. A pressure spike can cause a fitting to weep, disturb a sensitive sample, flex soft tubing, trigger a sensor alarm, or make a check valve chatter. It may last only a few milliseconds, so a slow data-acquisition setup can miss it completely. This article develops a

…